Preparing and Using the Automated Sample Changer (The 6-Pack)
1. Most Important Step:
Remove the existing sample in the magnet, type ej. Turn
the
lift air off by typing ij.
2. Turn on the 6-Pack air controller (on top of console) and cut the
air dryer off.
3. Remove any samples in the 6-Pack Holder.
4. Remove the prism and place on top of 6-Pack air controller (on top
of console).
5. Slide the 6-Pack Holder back and load sample (make sure depth
is set for each sample) with #1 at the front and #6 at the back
towards
the wall.
6. Make sure XWIN and ICON are running.
7. Click on the Automation icon.
8. Double click on the sample number you want run and fill in the
proper information.
9. You can change sweep widths and number of scans under the menu
option Parameters.
10. Click the Add button (bottom right) to perform an addition
experiment on the a sample.
11 When finished highlight ALL experiments and click Submit (lower
left)
12. Click on GO (upper left).
13. Just click Ok if you followed these direction and NO sample
is the machine.
14. When finished remove all samples from the holder.
15. Replace prism with small flat side pointing to the front.
16. Pull holder up against prism.
17. In XWIN type ej.
28. Insert the standard sample and turn OFF the 6-Pack air controller
(on top of console).
19. Type ij to insert the standard and turn the air dryer back
on.
20. Close your XWIN NMR session and maximize the existing XWIN session.
21. Your data can be found using File->Search->advanced->"given
filename"->"experiment #"